OmniaLab UP

Posted in Tap to Type II

OmniaLabUP
The Constant One
For pure H2O type ASTM II / CLRW (CLSI) / ISO 3696

Like its “little brother” OmniaTap TII, the OmniaLab dispenses Type II pure water at 2 l/min.

The OmniaLab is distinctly more powerful with pure water type ASTM II, where the pure water generation is up to 40 l/h. Furthermore, OmniaLab continually stores up to 100 litres of pure water type ASTM II / CLRW in the storage tank ready for dispensing. The flexible positioning of the purifying modules enables OmniaLab to be used as a mobile tower unit on castors, or space-saving model and can be stowed away in a laboratory base cabinet.

OmniaLab produces feedwater for autoclaves, clinical biochemistry analyzers or laboratory washing machines due to its pressurized tank outlet. Alternatively the front system dispensing fulfils demanding applications in scientific analysis such as tablet dissolution, buffer preparation, etc.

Features:

OptiFill dispenser is standard
The dual pure water cartridges provides reliable and long term Type II water supply
Microbial purity is ensured by interval recirculation and/or UV in the system.
Install as tower unit or under-table option
Type ASTM II dispense @ 2 l/min
Type ASTM II generation rate up to 40 l/h
100 litre storage tank for ASTM II water
Leakage sensor is standard
Unique non contacting level sensor helps keep water clean and fresh

Specifications:

 Specifications    OmniaLab  UP20    OmniaLab  UP40  
 Type II Water            
 Generation  performance @ 15°C    20    40  
  Conductivity   μS/cm    0.067 – 0.1    0.067 – 0.1  
 Resistivity  MΩ.cm    15 – 10    15 – 10  
 TOC  Value ppb    < 30    < 30  
 Dispensing  Rate l/min    Upto 2    Upto 2  
 Pure  water Pressurized outlet    Standard    Standard  
 Retention  of Bacteria, Particles and Pyrogens by RO Stage    95 – 99%    95 – 99%  
  Particles   0.2  μm/ml (Dispense from system dispenser)    No particles with size > 0.2 μm    No particles with size > 0.2 μm  
 Bacterial  Content** CFU/ml    < 0.01    < 0.01  
 UV Photo-Oxidation  (254nm)    Optional    Optional  
 Feedwater Specifications            
 Conductivity    < 2500    < 2500  
 TOC    < 2000    < 2000  
 Chlorine  content mg/l    < 5    < 5  
 Silt  Density / Fouling Index (SDI or FI)#    < 12##    < 12##  
 Manganese  and Iron content mg/l#    < 0.1    < 0.1  
 Temperature  °C    2 – 35    2 –  35  
 Technical Data            
 Operating  Pressure bar    2 – 6    2 – 6  
 Supply  Voltage Volts/Hz    90 – 240/50-60    90 – 240/50-60  
 Power  Consumption kW    0.1    0.1  
 Water  Connections    8mm    8mm  
 Ambient  Temperature    2 – 35    2 – 35  
 Dimensions  W X H X D mm    511 X 1520 X 575    511 X 1520 X 575  
 Weight    40    40  


* Dependent on feedwater quality       
**With sterilizing filter 0.2 μm and UV254nm


System Variants:

 Part No.    System  Name    Typical  Applications  
18600020   OmniaLabUP 20    Dissolutions apparatus, media and buffers  preparation, glassware rinsing, microbiology, feedwater to clinical  biochemistry Analyzers, OmniaPure systems, etc…  
18600020 OmniaLab UP 40


Accessories:

 Part No.    System  Name  
 19200021#    3  Stage Automatic Prefiltration 10”  
 18200023    3  Stage Prefiltration 20”  
 19200050    UV  Tank disinfection unit 254nm  
 19200051    UV  flow through disinfection 254nm  
 19200100    Tank  100 ltr  
 16125000##    Mix  Multi Mini pretreatment for SDI above 12  
 19102100    POU  Biofilter